Installation Note N9020-90181
7
Option LNP or Option MPB Installation Kit
Manually Control the Option
Assure the signal analyzer is in Spectrum Analyzer mode.
View the 4.8 GHz calibrator signal:
Preset the analyzer by pressing the green
Mode Preset
key.
Press
Input/Output
,
RF Calibrator
,
4.8 GHz
.
Press
Freq/Channel
, enter
4.8 GHz
. Press
Span
, enter
5 MHz
.
Enable the Low Noise Path or the Preselector Bypass:
Press
AMPTD
,
More
,
uW Path Ctrl
.
Softkeys will appear that let you select Low Noise Path Enable or uW Preselector Bypass.
Pressing either of these softkeys will result in an audible “click” that is caused by mechanical
switches activating. The 4.8 GHz calibrator signal will remain on screen.