Installation Note N9020-90181

3

Option LNP or Option MPB Installation Kit

Option LNP or Option MPB Installation Kit 

Introduction

When a new instrument is ordered from the factory with either Option LNP or MPB enabled, the 

instrument is assembled with hardware that supports both options. Therefore, instruments that 

currently contain Option LNP can be upgraded to also contain Option MPB by licensing Option 

MPB. Instruments that currently contain Option MPB can be upgraded to also contain Option LNP 

by licensing Option LNP. 
This installation note explains how to install Option LNP or Option MPB into instruments that 

already contain one of these options.
If the instrument HAS frequency range option 508, 513, or 526 and DOES NOT currently contain 

either Option LNP, or Option MPB, the hardware required to support these two options is not 

contained in the instrument and the N9030AK-HL1 retrofit kit is required.
If the instrument HAS frequency range option 543, 544, or 550 and DOES NOT currently contain 

either Option LNP, or Option MPB, the hardware required to support these two options is not 

contained in the instrument and the N9030AK-HL2 retrofit kit is required.
Option LNP provides a switchable Low Noise Path, for input frequencies 3.6 GHz and greater. This 

low noise path when activated allows enhanced DANL and SHI performance. 
Option MPB provides a switchable microwave preselector filter bypass path for input frequencies 

3.6 GHz and above. The preselector filter limits the bandwidth of the input signal and can cause 

slight amplitude errors if the filter drifts.

Products Affected:

PXA N9030A, Options 508, 513, 526, 543, 544, and 550

Serial Numbers: 

All

To Be Performed By:  

(X) Keysight Service Center 
(X) Personnel Qualified by Keysight
(X) Customer

Estimated Installation Time:  

Estimated Adjustment and Verification Time:

0.5 Hours

4 Hours (see Note 2)

Summary of Contents for PXA N9030A

Page 1: ...Preselector Bypass Keysight PXA Signal Analyzers Installation Note Notice This document contains references to Agilent Please note that Agilent s Test and Measurement business has become Keysight Technologies For more information go to www keysight com ...

Page 2: ...ited to the implied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Manual Part Number N9020 90181 Edition Edition 1 May 2015 Printed in USA Malaysia Published by Keysight Technologies Inc 1400 Foun...

Page 3: ... options is not contained in the instrument and the N9030AK HL1 retrofit kit is required If the instrument HAS frequency range option 543 544 or 550 and DOES NOT currently contain either Option LNP or Option MPB the hardware required to support these two options is not contained in the instrument and the N9030AK HL2 retrofit kit is required Option LNP provides a switchable Low Noise Path for input...

Page 4: ...options If you require test data for these options then a full calibration is required after installation of these options Arrangements regarding the level of calibration must be made with the calibration provider 2 If this instrument has been adjusted as part of a repair or calibration during its first year or if the instrument is more than one year old additional adjustments and performance veri...

Page 5: ...Certificate 5964 5178 from the kit 2 Redeem the Option Upgrade Entitlement Certificate by following the instructions on the Certificate 3 After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an attached License File 4 Locate a USB storage device Perform a virus scan on this device before use 5 Save the License File to the root directory of the USB storage devi...

Page 6: ...Installation message as shown in Figure 2 Figure 2 Successful License Installation Alternate Installation Procedure The License File can be manually installed over USB or LAN by placing the license file in the following folder on the signal analyzer C Program Files Agilent licensing Verify the Installation 1 Cycle the instrument power and allow the instrument to boot to the measurement screen 2 Pr...

Page 7: ...reset key Press Input Output RF Calibrator 4 8 GHz Press Freq Channel enter 4 8 GHz Press Span enter 5 MHz Enable the Low Noise Path or the Preselector Bypass Press AMPTD More uW Path Ctrl Softkeys will appear that let you select Low Noise Path Enable or uW Preselector Bypass Pressing either of these softkeys will result in an audible click that is caused by mechanical switches activating The 4 8 ...

Page 8: ... the following performance verification tests are the minimum set required to ensure that this newly installed option is functioning properly Performing only these tests does not guarantee the instrument meets all specifications A full calibration is required to assure the instrument meets all specifications The end user must ultimately determine whether they want a full calibration to be performe...

Page 9: ...This information is subject to change without notice Keysight Technologies 2012 2015 Edition 1 May 2015 www keysight com ...

Reviews: