In This Guide
Safety, Regulatory, and Environmental Information
Safety Information
Complies with the essential requirements of the European Low Voltage
Directive as well as current editions of the following standards (dates and
editions are cited in the Declaration of Conformity):
— IEC/EN 61010-1
— Canada: CAN/CSA-C22.2 No. 61010-1
— USA: UL std no. 61010-1
Environmental Information
Samples of this product have been type-tested in accordance with the
Keysight Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation and end-use; those
stresses include but are not limited to temperature, humidity, shock,
vibration, altitude, and power-line conditions. Test methods are aligned
with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.
Parameter
Required Values/Ranges
Operating Environment
For Indoor Use ONLY
This instrument has no air filters on the fan ports. Operation in dirty,
dusty, or similar environments must be avoided.
Operating Temperature
0 °C to +40 °C
The instrument powers-up and displays no error messages within this
temperature range (except for "source unleveled" error message that may
occur at temperatures outside the specified performance temperature
range of 25 ± 5 °C).
Storage Temperature
–40 °C to +70 °C
Error-corrected range
23 °C ±3 °C with less than 1 °C deviation from calibration temperature.