Keysight N5990A MIPI M-PHY Receiver Test User Guide
283
Sys Burst Tests
8
Results
An example of a HTML Viewer for the Test 2.3.5-SYS-RX Differential
Termination Disable Time (T
TERM-OFF-SYS-RX
) test procedure is shown in
. The results comprise of the following:
• The common parameter list (in
Expert Mode
).
• A test data table (Refer to
Figure 163
Example HTML Viewer for Test 2.3.5-SYS-RX Differential Termination Dis-
able Time (T
TERM-OFF-SYS_RX
) Test
Table 52
Test 2.3.5-SYS-RX Differential Termination Disable Time (T
TERM-OFF-SYS-RX
) data table
Parameter Name
Parameter Description
Result
"Pass"/"Fail"- If the value is "Fail", the termination disable time is not within the spec limit.
Termination Time Min
It is the shortest/minimum termination disable time in "ns".
Termination UI Min
It is the shortest/minimum termination disable time in "UI".
Min Spec
It is the minimum value of the termination disable time, where the DUT must work according to the
specification.