3-10 Keysight N5290/1A Service Guide
System Tests and Adjustments
The N5290/1A Millimeter Test System Preliminary Check
3-
2.
The Millimeter Configuration window that opens should have N5290A
Broadband or N5291A Broadband highlighted.
Else, press N5290A Broadband or N5291A Broadband. Refer to
3.
Press Installation Cal. In the dialog box that opens, connect a load and
press
Measure
.
Figure 3-3
Millimeter Configuration Window
Figure 3-4
Millimeter Installation Cal – Dialog Box
For best results use a load that ex tends to 50 GHz.