N2893A Probe
N2893A 100 MHz Current Probe User’s Guide
21
Characteristics and Specifications
Table 2
Measurement Characteristics
Item
Characteristic
a
a Requires 1M
Ω
termination. Valid for 23°C ±3°C (73°F ±5°F), at least 30 minutes after power on.
Probe Bandwidth
DC to 100 MHz (–3 dB)
Accuracy (Probe Only)
b
b This is a specification and is guaranteed at 23°C ±3°C (73°F ±5°F).
±1% of reading ±1mV (DC or 45 Hz to 66 Hz)
Risetime
c
c Rise time is calculated as: Tr = 0.35/Bandwidth.
~
3.5 ns
Maximum Continuous Current
15A peak, 15 A
DC
, 10 A
rms
30A peak, 30 A
DC
, 24 A
rms
(when one probe is used with InfiniiVision
3000XT, 5000/6000/7000 scope)
Maximum Non-continuous Peak
Current (for Pulse Widths
~
10µs)
50A
30A peak,
32A peak (when one probe is used with InfiniiVision 3000XT,
5000/6000/7000 scope)
Probe Sensitivity
0.1V/A
Noise
~
2.5 mA
rms
(with 20 MHz Bandwidth limiting)
Insertion Impedance
See
Figure 12
on page 25.
Temperature Coefficient
~
±2% (0
°
C to 40
°
C)
Maximum Circuit Voltage
300V (refer to safety considerations and product compliances)
Not rated for CAT II, III, IV