18
N2893A 100 MHz Current Probe User’s Guide
N2893A Probe
Figure 7
Providing power to the N2893A probe
2
Turn on the equipment.
3
Wait for 20 minutes to allow the probe to warm up.
4
Connect the test setup as shown in
Figure 8
on page 19 as per the following substeps.
a
Clamp the N2893A probe around the test cable <1> and lock the sensor head of the
probe.
b
Press the
DEMAG
button on the probe.
c
Set the digital multimeter (DMM <1>) to A
rms
mode.
d
Connect the test cable <1> to the positive terminal of the AC current generator's
OUTPUT terminal block and to the current input of the DMM <1>.
e
Connect the test cable <2> to the negative terminal of the AC current generator's
OUTPUT terminal block and to negative input of the DMM <1>.
f
Using the DMM <1>, set the output current of AC current generator to 6A
rms
.