11
Functional Tests
Test Equipment
Test Equipment
The table below summarizes the test equipment needed to perform all of the
functional tests. Alternate equipment model numbers are given in case the
recommended equipment is not available. If neither the recommended nor the
alternative test equipment are available, substitute equipment that meets or
exceeds the critical specifications listed.
Table 1-1
Analyzer
Option
Item
Critical Specifications
Recommended
Keysight Model
Alternate
Keysight
Model
Adapters
All
3.5 mm (f) to 3.5mm (f)
(connector saver for source)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
83059B
1250-1749
All
BNC (f) to SMA (m)
Frequency: 40 MHz
1250-1200
All
Type N (f) to Type N (f)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.05:1
1250-1472
1250-0777
All
Type N (m) to 3.5 mm (m)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1743
All
Type N (m) to 3.5 mm (f)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1744
All
Type N (f) to 3.5 mm (f)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1745
All
Type N (m) to BNC (f)
Frequency: 10 MHz to 1 GHz
VSWR: < 1.05:1
1250-1476
Attenuators
All
10 dB Step Attenuator
Frequency: 50 MHz
Range
: 0 to 70 dB
8495A
8496A
All
10 dB Fixed Attenuator (2
required)
Frequency: 50 MHz
VSWR: < 1.2:1
8493C Option 010
8493A Option
010
or
8493B Option
010
Pxx
30 dB Fixed Attenuator
Accuracy
: < 0.05 dB @ 50 MHz
11708A
Cables
All
3.5 mm (1 meter)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Loss: < 2.0 dB
11500E
8120-4921
All
Cable, BNC (3 required)
120 cm (48 in.) BNC cable
10503A