15
Keysight N9020A MXA Signal Analyzer
Functional Tests Guide
2
Displayed Average Noise Level (DANL)
Test Limits (with 0 dB input attenuation)
See
for values.
The Displayed Average Noise Level (DANL) of the signal analyzer is measured
across a 10 kHz frequency span at several center frequencies. The analyzer
input is terminated into a 50
Ω
load. A test is performed to assure the
measurement is not performed in the presence of a residual response. The
measurement is then averaged, and the result is normalized to a 1 Hz
bandwidth.
Figure 2-1
DANL Test Setup
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Termination, 50
Ω
Type-N(m)
Frequency: DC to 18 GHz
909A Option 012