M9506A-BC1 5-Slot AXIe Chassis with Integrated Controller User Guide
61
SCPI Programming Interface
4
:DIGitizer[N]:RESet
Command
:DIGitizer[N]:RESet
Description
This command resets digitizer [N] to defaults.
Example
-> dig3:res
:DIGitizer[N]:SCALibration[?]
Command
:DIGitizer[N]:SCALibration
Description
This command initiates the Self-Calibration of digitizer[N]. No other arguments are
required.
Example
-> dig3:scal
Query
:DIGitizer[N]:SCALibration?
Description
This query returns the status of digitizer self-calibration, 1 if not complete, 0 if
complete.
Query Example
-> dig3:scal?
<- 0
:DIGitizer[N]:SCALibration:REQuired?
Query
:DIGitizer[N]:SCALibration:REQuired?
Description
This query returns whether the digitizer self-calibration is required or not, 1 if
self-calibration is required, 0 if not.
Self-calibration should be performed after the module has warmed up to a stable
operating temperature, approximately 20 minutes. Self-calibration only calibrates
the digitizer for the current settings and will need to be performed when the
acquisition configuration changes. If the acquisition is attempted without
self-calibration, the digitizer will not arm, and the arm command execution will
return an error. Each digitizer needs to have self-calibration initiated separately,
but any number of digitizers may have self-calibration initiated at the same time.
Example
-> dig3:scal:req?
<- 0