1 Installing the PXI Chassis Hardware
Keysight
M9347A
H02
Getting
Started
Guide
3
Set Up the PXI Chassis
1.
Using
the
PXI
Chassis
’
Installation
and
Quick
Start
Guide
https://literature.cdn.keysight.com/litweb/pdf/M9019-90001.pdf?id=2766869
set
up
the
PXI
Chassis
.
Do
not
power
on
the
chassis.
2.
Insert
the
PXI
modular
instrument
cards
and
controller
into
the
appropriate
slots
according
to
the
instillation
manual.
3.
Make
sure
all
cards
are
securely
locked
in
place.
4.
Power
on
the
PXI
chassis.
Do not remove or insert any PXI cards into the chassis with the
power on. This could cause irreparable damage to both the cards
and the chassis.
Ventilation Requirements: When installing the product in a
cabinet, the convection into and out of the product must not be
restricted. The ambient temperature (outside the cabinet) must
be less than the maximum operating temperature of the
instrument by 4
o
C for every 100 watts dissipated in the cabinet. If
the total power dissipated in the cabinet is greater than 800
watts, then forced convection must be used.
Consider this when selecting the location of your system.
Hookup Guide
Do
not
turn
on
the
M9347A
without
a
clock
signal
input.
See
"Clock Selection" on page 8.
The ideal clock drive level is between 0 and -6 dBm.
Contacting
Keysight
Assistance
with
test
and
measurements
needs
and
information
on
finding
a
local
Keysight
office
are
available
on
the
Web
at:
http://www.keysight.com/find/assist