Commands Quick Reference
5
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
77
Table 4
:ACQuire Commands Summary
Command
Query
Options and Query Returns
n/a
:ACQuire:AALias? (see
{1 | 0}
:ACQuire:COMPlete
<complete> (see
:ACQuire:COMPlete?
<complete> ::= 100; an integer in
NR1 format
:ACQuire:COUNt
<count> (see
:ACQuire:COUNt? (see
<count> ::= an integer from 2 to
65536 in NR1 format
:ACQuire:DAALias
<mode> (see
:ACQuire:DAALias?
<mode> ::= {DISable | AUTO}
:ACQuire:DIGitizer
{{0 | OFF} | {1 |
ON}} (see
:ACQuire:DIGitizer?
{0 | 1}
:ACQuire:MODE <mode>
(see
:ACQuire:MODE? (see
<mode> ::= {RTIMe | ETIMe |
SEGMented}
:ACQuire:POINts[:ANAL
og] <points> (see
:ACQuire:POINts[:ANAL
<points> ::= {AUTO |
<points_value>}
<points_value> ::= desired
analog memory depth in integer
NR1 format
:ACQuire:POINts[:ANAL
og]:AUTO {{0 | OFF} |
{1 | ON}} (see
:ACQuire:POINts[:ANAL
og]:AUTO? (see
{0 | 1}
:ACQuire:RSIGnal
<ref_signal_mode>
(see
:ACQuire:RSIGnal?
<ref_signal_mode> ::= {OFF | OUT
| IN}
:ACQuire:SEGMented:AN
ALyze (see
n/a
n/a
:ACQuire:SEGMented:CO
UNt <count> (see
:ACQuire:SEGMented:CO
<count> ::= an integer from 2 to
1000 in NR1 format
:ACQuire:SEGMented:IN
Dex <index> (see
:ACQuire:SEGMented:IN
<index> ::= an integer from 1 to
1000 in NR1 format
:ACQuire:SRATe[:ANALo
g] <rate> (see
:ACQuire:SRATe[:ANALo
<rate> ::= {AUTO | <sample_rate>}
<sample_rate> ::= desired analog
sample rate in NR3 format