Keysight Technologies FieldFox N9116A Security Features And Volatility Documentation Download Page 9

 9

Keysight Technologies FieldFox Handheld Analzers 
N991xA, N992xA, and N993xA

Security Features and Volatility Documentation

3

Terms and Definitions

Definitions

Clearing

 - Clearing is the process of eradicating the data on media before 

reusing the media so that the data can no longer be retrieved using the 

standard interfaces on the instrument.  Clearing is typically used when the 

instrument is to remain in an environment with an acceptable level of 

protection.

Sanitization

 - Sanitization is the process of removing or eradicating stored 

data so that the data cannot be recovered using any known technology. 

Instrument sanitization is typically required when an instrument is moved from 

a secure to a non-secure environment such as when it is returned to the factory 

for calibration. (The instrument is declassified)  Keysight memory sanitization 

procedures are designed for customers who need to meet the requirements 

specified by the US Defense Security Service (DSS). These requirements are 

outlined in the "Clearing and Sanitization Matrix" issued by the Cognizant 

Security Agency (CSA) and referenced in National Industrial Security Program 

Operating Manual  (NISPOM) DoD 5220.22M ISL 01L-1 section 8-301.

Security erase

 - Security erase is a term that is used to refer to either the 

clearing or sanitization features of Keysight instruments. 

Instrument declassification

 - A term that refers to procedures that must be 

undertaken before an instrument can be removed from a secure environment 

such as is the case when the instrument is returned for calibration. 

Declassification procedures will include memory sanitization and or memory 

removal. Keysight declassification procedures are designed to meet the 

requirements specified by the DSS NISPOM security document (DoD 5220.22M 

chapter 8).

Summary of Contents for FieldFox N9116A

Page 1: ...Features and Volatility Documentation Keysight Technologies FieldFox Handheld Analyzer N9912A N9913A N9914A N9915A N9116A N9917A N9918A N9923A N9923AN N9925A N9926A N9927A N9928A N9935A N9936A N9937A...

Page 2: ...Software to U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula T...

Page 3: ...s Definitions 9 4 System Components 5 Instrument Memory and Volatility Information 6 Memory Clearing Sanitization and or Removal Procedures 7 User and Remote Interface Security Measures Screen and Ann...

Page 4: ...4 Contents...

Page 5: ...formation on finding a local Keysight office is available on the internet at http www keysight com find assist If you do not have access to the internet please contact your field engineer In any corre...

Page 6: ...6 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Contact Keysight Sales and Service Offices...

Page 7: ...the security instrument tab Service Guide N9912A N9912 90003 http cp literature keysight com litweb pdf N9912 90003 pdf Service Guide N9923A N9923AN N9923 90018 http cp literature keysight com litweb...

Page 8: ...8 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Product Declassification and Security Introduction...

Page 9: ...nstrument is declassified Keysight memory sanitization procedures are designed for customers who need to meet the requirements specified by the US Defense Security Service DSS These requirements are o...

Page 10: ...10 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Terms and Definitions Definitions...

Page 11: ...tem board and RF board N9916A FieldFox Handheld Analyzer Vector Network Analyzer and Spectrum Analyzer Volatile and Non volatile memory on SOM board system board and RF board N9917A FieldFox Handheld...

Page 12: ...le and Non volatile memory on SOM board system board and RF board N9936A FieldFox Handheld Analyzer Spectrum Analyzer Volatile and Non volatile memory on SOM board system board and RF board N9937A Fie...

Page 13: ...a Volatile and Non volatile memory on SOM board system board and RF board SOM board Divided into three partitions Keysight 42 Mb FactoryData 2 Mb and UserData 20 Mb Only the UserData partition is acce...

Page 14: ...tting System board and RF board Each PCA has an ID EPROM that stores PCA identification and revision information Not writable by the user Flash memory 16 Kb and volatile memory 2 Kb No No Yes Front pa...

Page 15: ...64 MB Memory clearing Memory Erase function Memory sanitization The following process completely clears all user accessible memory on the FieldFox instrument This process requires firmware revision A...

Page 16: ...from the partition When the instrument re boots the firmware writes default versions of system files that are missing Thus clearing the UserData partition does not permanently interfere with normal o...

Page 17: ...the reboot is completed the instrument will be ready for normal operation Memory sanitization The following process completely clears all user accessible memory on the FieldFox instrument This process...

Page 18: ...ystem Board Micro Controller 32 Kb Flash Memory 1 Kb Volatile Memory Description and purpose For System Board System Monitor Micor Controller Size 32 Kb Flash Memory 1 Kb Volatile Memory Memory cleari...

Page 19: ...mory sanitization N A Memoryremoval This memory can not be removed without damaging the instrument Write protecting N A Memory validation N A Remarks Table 6 7 Flash Memory on Main Battery 1 Kb Descri...

Page 20: ...20 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Memory Clearing Sanitization and or Removal Procedures...

Page 21: ...rol removable USB SD card mass storage permissions from the FieldFox user interface at this time The customer is responsible for managing removable storage media in secure locations How to set Securit...

Page 22: ...ation Blanking Remote Access Interfaces The user is responsible for providing security for the LAN port for remote access by controlling physical access to the LAN port The LAN port must be controlled...

Page 23: ...e new board even though the instrument is still under warranty The 4GB microSD resides in System Board If the instrument is not functioning and you are unable to use the security functions to clear th...

Page 24: ...24 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Procedure for Declassifying a Faulty Instrument...

Reviews: