Memory Clearing, Sanitization and/or Removal Procedures

Table 6-2 

4 GB Internal microSD Card

Description and 

purpose

Stores Operating System, Application Firmware and some User Data.  Only the 

User Data is accessible by the user
 

Size

4 GB

Memory clearing 

The following process completely clears all user accessible data on the 

FieldFox instrument. This process requires firmware revision A.03.02 or later. 
Press the following keys:

1.

System(7) 

2.

Service Diagnostics

3.

Advanced

4.

Erase User Data

5.

Confirm Erase

Wait for the instrument to reboot and begin normal operation.
The process will completely erase the UserData on the 4GB Internal microSD 

Card and reboot the instrument.  When the reboot is completed, the 

instrument will be ready for normal operation.

Memory 

sanitization 

The following process completely clears all user accessible memory on the 

FieldFox instrument. This process requires firmware revision A.03.02 or later.
Press the following keys:

1.

System(7)

2.

Service Diagnostics

3.

Advanced 

4.

Erase User Data

5.

Confirm Erase

Wait for the instrument to reboot and begin normal operation.
The process will completely erase the Data on the 4GB Internal microSD Card 

and reboot the instrument.  When the reboot is completed, the instrument will 

be ready for normal operation.

Memory removal           

 

This memory can not be removed without damaging the instrument

Write protecting

N/A

Memory 

validation

N/A

Remarks

4GB microSD is only used in FieldFox model N9913A, N9914A, N9915A, 

N9916A, N9917A, N9918A, N9925A, N9926A, N9927A, N9928A, N9935A, 

N9936A, N9937A, N9938A 
UserData on internal flash (64 MB flash memory) is not available when 

microSD flash memory is used. 

Summary of Contents for FieldFox N9116A

Page 1: ...Features and Volatility Documentation Keysight Technologies FieldFox Handheld Analyzer N9912A N9913A N9914A N9915A N9116A N9917A N9918A N9923A N9923AN N9925A N9926A N9927A N9928A N9935A N9936A N9937A...

Page 2: ...Software to U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula T...

Page 3: ...s Definitions 9 4 System Components 5 Instrument Memory and Volatility Information 6 Memory Clearing Sanitization and or Removal Procedures 7 User and Remote Interface Security Measures Screen and Ann...

Page 4: ...4 Contents...

Page 5: ...formation on finding a local Keysight office is available on the internet at http www keysight com find assist If you do not have access to the internet please contact your field engineer In any corre...

Page 6: ...6 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Contact Keysight Sales and Service Offices...

Page 7: ...the security instrument tab Service Guide N9912A N9912 90003 http cp literature keysight com litweb pdf N9912 90003 pdf Service Guide N9923A N9923AN N9923 90018 http cp literature keysight com litweb...

Page 8: ...8 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Product Declassification and Security Introduction...

Page 9: ...nstrument is declassified Keysight memory sanitization procedures are designed for customers who need to meet the requirements specified by the US Defense Security Service DSS These requirements are o...

Page 10: ...10 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Terms and Definitions Definitions...

Page 11: ...tem board and RF board N9916A FieldFox Handheld Analyzer Vector Network Analyzer and Spectrum Analyzer Volatile and Non volatile memory on SOM board system board and RF board N9917A FieldFox Handheld...

Page 12: ...le and Non volatile memory on SOM board system board and RF board N9936A FieldFox Handheld Analyzer Spectrum Analyzer Volatile and Non volatile memory on SOM board system board and RF board N9937A Fie...

Page 13: ...a Volatile and Non volatile memory on SOM board system board and RF board SOM board Divided into three partitions Keysight 42 Mb FactoryData 2 Mb and UserData 20 Mb Only the UserData partition is acce...

Page 14: ...tting System board and RF board Each PCA has an ID EPROM that stores PCA identification and revision information Not writable by the user Flash memory 16 Kb and volatile memory 2 Kb No No Yes Front pa...

Page 15: ...64 MB Memory clearing Memory Erase function Memory sanitization The following process completely clears all user accessible memory on the FieldFox instrument This process requires firmware revision A...

Page 16: ...from the partition When the instrument re boots the firmware writes default versions of system files that are missing Thus clearing the UserData partition does not permanently interfere with normal o...

Page 17: ...the reboot is completed the instrument will be ready for normal operation Memory sanitization The following process completely clears all user accessible memory on the FieldFox instrument This process...

Page 18: ...ystem Board Micro Controller 32 Kb Flash Memory 1 Kb Volatile Memory Description and purpose For System Board System Monitor Micor Controller Size 32 Kb Flash Memory 1 Kb Volatile Memory Memory cleari...

Page 19: ...mory sanitization N A Memoryremoval This memory can not be removed without damaging the instrument Write protecting N A Memory validation N A Remarks Table 6 7 Flash Memory on Main Battery 1 Kb Descri...

Page 20: ...20 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Memory Clearing Sanitization and or Removal Procedures...

Page 21: ...rol removable USB SD card mass storage permissions from the FieldFox user interface at this time The customer is responsible for managing removable storage media in secure locations How to set Securit...

Page 22: ...ation Blanking Remote Access Interfaces The user is responsible for providing security for the LAN port for remote access by controlling physical access to the LAN port The LAN port must be controlled...

Page 23: ...e new board even though the instrument is still under warranty The 4GB microSD resides in System Board If the instrument is not functioning and you are unable to use the security functions to clear th...

Page 24: ...24 Keysight N991xA N992xA N993xA Security Features and Volatility Documentation Procedure for Declassifying a Faulty Instrument...

Reviews: