Keysight Technologies E5400-Pro Series Soft Touch Connectorless Probes User’s Guide
4
Operating the E5402A, E5405B, and E5406A-Pro Series Probes
Equivalent Probe Loads
The following probe load models are based on in-circuit measurements
made with a Keysight 8753E 6 GHz network analyzer and a Keysight
54750A TDR/TDT using a 50
Ω
test fixture. The following schematic
accurately models the probe load out to 6 GHz. The figure on the following
page shows the agreement between measured impedance and this model.
PC board pads are not included.
Figure 28 Probe load model (E5402A, E5405B, and E5406A)
20K
Ω
Rtip1
1.17nH
L12
0.63nH
L11
C12
0.280pF
Rgnd1
0.5
Ω
Cm12
0.070pF
Rtrm1
75
Ω
.350pF
Cshnt1
20K
Ω
Rtip2
1.17nH
L22
0.63nH
L21
C22
0.280pF
Rgnd2
0.5
Ω
Rtrm2
75
Ω
.350pF
Cshnt2
D1
D0
+0.75V
+0.75V