Chapter 4
67
Troubleshooting
Performance test failure troubleshooting
4. T
roub
lesho
o
ting
Performance test failure troubleshooting
This section describes the adjustment and troubleshooting procedures used when the
E4981A fails the performance tests. If the performance of the instrument is critical for the
test limits and seems to be adjustable, perform first the adjustment(s) related to the failed
test. When the test result is far from the tolerance of the test or the performance is not
adjustable, isolate the faulty assembly in accordance with the "Performance tests failure
troubleshooting procedure".
Performance test failure troubleshooting
represents the relationships between the failed test and probable faulty
assembly. If the performance test failure cannot be removed by a proper adjustment,
replace the assembly shown in this table.
Note that this table lists some typical cases. There are possibilities that another assembly
may be faulty.
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
Table 4-11
Performance test failure troubleshooting information
Test
Num.
First failed test
Probable faulty board
assembly
A31
A34
PPMC
1
Test Signal Frequency Accuracy Test
#
###
2
Test Signal Level Accuracy Test
#
###
3
Impedance Measurement Accuracy Test
#
###
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...