![Keysight Technologies E36100B Series Operating And Service Manual Download Page 83](http://html1.mh-extra.com/html/keysight-technologies/e36100b-series/e36100b-series_operating-and-service-manual_1962204083.webp)
Test Description
Instrument settings
Voltage Programming & Readback, Zero Voltage (V0)
0 V
1 A
Voltage Programming & Readback, Maximum Voltage (Vmax)
35 V
1 A
CV Load Regulation, Line Regulation, Ripple and Noise
35 V
1 A
Transient Response
35 V
0.5 A to 1 A
Current Programming & Readback, Zero Current (I0)
0 A
35 V
Current Programming & Readback, Maximum Current (Imax)
1 A
35 V
CC Load Regulation, Line Regulation, Ripple and Noise
1 A
35 V
Keysight E36100B Series Operating and Service Guide