Standard Event Status Enable Register
Bit Value Name
Description
0
1
OPC
Operation complete
1
2
(unused) (reserved for future use)
2
4
QYE
Query error
3
8
DDE
Device dependent error
4
16
EXE
Execution error
5
32
CME
Command error
6
64
(unused) (reserved for future use)
7
128
PON
Power-on
Operation Status Register
Bit Value Name
Description
0
1
CAL
The output is computing new calibration constants
1-4 2-16 (unused) (reserved for future use)
5
32
WTG
The output is waiting for a trigger.
6-7 64-128 (unused) (reserved for future use)
8
256
CV
The output is in constant voltage mode.
9
512
(unused) (reserved for future use)
10
1024
CC
The output is in constant current mode.
Questionable Register
Bit Value Name
Description
0
1
OV
Output disabled by overvoltage protection
1
2
OC
Output disabled by overcurrent protection
2-3
4-8
(unused) (reserved for future use)
4
16
OT
Output disabled by overtemperature protection
5-9 32-512 (unused) (reserved for future use)
10
1024
UNR
Output is unregulated