Programming Ranges
The following table shows the maximum voltage, current, and power that can be programmed for each model. The
DEFault voltage is always 0 V.
Model
Max voltage (V) Max current (A) Max power (W)
E36102B
6
5.0
30
E36103B
20
2.0
40
E36104B
35
1.0
35
E36105B
60
0.6
36
E36106B
100
0.4
40
Note that the actual programmable values are 3% above these nominal values, but instrument specifications are
only guaranteed up to the values shown in this table.
Table of programming and readback resolutions
Model
Voltage programming Voltage readback Current programming Current readback Small current readback
E36102B
0.001 V
0.001 V
0.001 A
0.001 A
0.001 mA
E36103B
0.001 V
0.001 V
0.001 A
0.001 A
0.001 mA
E36104B
0.001 V
0.001 V
0.001 A
0.001 A
0.001 mA
E36105B
0.001 V
0.01 V
0.001 A
0.0001 A
0.001 mA
E36106B
0.001 V
0.01 V
0.001 A
0.0001 A
0.001 mA
Keysight E36100B Series Operating and Service Guide