N9069C Noise Figure Measurement Guide 23
Making Noise Figure Measurements
Making Frequency Converter Measurements
Initial Setup Procedure
Follow the overview procedure of the initial setup.
Diagram
Calibration
or
Measurement
. This does not affect the
measurement or calibration, but indicates how the noise source,
the DUT and the analyzer should be set up. The blue ‘eye’ icon
acts as a visual reminder of the
Frequency Context
setting
you have selected.
DUT Input
(RF)
Start: 3.2 GHz
DUT Input
(RF)
Stop:3.7 GHz
DUT Out (IF)
Fixed: 60 MHz
Table 2-2
DUT Setup Table
Step
Action
Notes
1.
Turn the instrument on
— Press the power key
To obtain optimum measurement accuracy, it is
recommended the analyzer warm up for 30
minutes.
2.
Select the mode and
measurement
— Press
MODE/MEAS
,
Noise Figure
,
OK
3.
Preset the mode
— Press
Mode Preset
4.
Configure the ENR
— Press
MEAS SETUP,
ENR.
— Toggle
ENR Mode
to
Table
— Toggle
Use Meas Table
Data for Cal
to
On.
5.
View ENR data
— Press
Edit Meas Table
to view the ENR values
automatically loaded from the
SNS noise source into the
instrument.
— Press
Close
.
If you are using other noise sources, for example,
the Keysight 346B noise source, you need to enter
the ENR values manually.
For more information about entering the ENR
data, refer to
.
Note that for calibration and for making
measurements, the ENR data for different
frequency ranges are used.