3) Press to edit the
Trigger type
, and then select to set the
Trigger type
to
TIMER
;
this will make a time domain measurement.
4) Rotate to select the
Trigger Parameters
for Channel 1 and configure them as
shown below. These parameters define the number of the measurement points and the
measurement interval.
(Measurement Trigger Count: 5000, Measurement Trigger Period: 10us)
5) Press repeatedly until
Graph View
is shown in the display.
6) Press to source the waveform and monitor the signal.
(The status information will display during the waveform sourcing.)
7) Press to adjust the graph scale after the waveform sourcing is complete.
You can now see the complete waveform on the B2961A/62A’s GUI as shown below.
13 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide