Objective

This lab shows how a single-point 

measurement can be done with 

an intuitive GUI and shallow menu 

tree on the B2900A Series of SMUs 

through the resistance measurement

Procedure overview

1. Change View mode to Single View

2. Use AUTO measurement operation

3. Turn off the channel output

LAB 1: Make a Resistance Measurement

Setup

1. Connect the yellow banana plug to Ch 1 Low Force terminal.

2. Connect the red banana plug to Ch 1 High Force terminal.

3. Clip the one lead of 1 kΩ Resistor with the black gold-plated tweezers.

4. Clip the other lead of 1 kΩ Resistor with the red gold-plated tweezers.

Demonstration

1. Change View mode to Single View

 

a. Press  

 repeatedly until Single View for 

  Channel 1 is shown in the display.

 

2. Use AUTO measurement operation

 

a. Rotate           to select Resistance measurement operation, and then press   
  to edit it. Then select            to set Resistance measurement operation to AUTO.

 

 

b. Press Ch1            to turn on Channel 1 Output relay.

 

c. Press            to perform a single point measurement.

4

1

2

3

Red (11059A)

Yellow (11059A)

Black (11059A)

Red (11059A)

Black

Brown

Brown

Brown

Brown

1 kΩ Resistor

(2) Press AUTO

(1) Rotate and press 

Resistance Measurement Operation

to select

Measurement parameters are updated whenever

Trigger Button is pressed

Summary of Contents for B2901A

Page 1: ...Keysight Technologies Detailed Demonstration Guide Making Resistance Measurement Using SMU Keysight B2901A 02A 11A 12A Precision Source Measure Unit Demonstration Guide...

Page 2: ...plication software that make it easy for you to begin making productive measure ments immediately This demonstration guide shows how easily basic resistance measurement can be made using the Keysight...

Page 3: ...ed tweezers 4 Clip the other lead of 1 k Resistor with the red gold plated tweezers Demonstration 1 Change View mode to Single View a Press repeatedly until Single View for Channel 1 is shown in the d...

Page 4: ...o stop making measurements periodically 3 Turn off the channel output a Press Ch1 to turn off Channel 1 Output relay LAB 1 Make a Resistance Measurement continued Measurement parameters are updated pe...

Page 5: ...Resistor with black gold plated tweezers Why is 4 wire connection required For very low resistance measurements where the residual lead resistance is comparable to the DUT resistance the 2 wire measu...

Page 6: ...wire connection you can see the status indicator on GUI as below although no indicator can be seen on being configured to use 2 wire connection Demonstration 1 Reset the instrument a Press and then pr...

Page 7: ...nd select and then press to configure to use 4 wire connection c Press then press to set Channel 1 V I Source Function to I Source If can t be found on the Assist keys press to change the keys d Press...

Page 8: ...asure condition and calculates the resistance value e Press and set Channel 1 Limit value to 1 V f Rotate to select Resistance measurement operation and press to edit it Then press to set Resistance m...

Page 9: ...connection is 1 6 Ohm The difference that is 0 6 Ohm should be the residual lead resistance on the measurement cables LAB 2 Make an Accurate Low Resistance Measurement continued 3 Press OK 2 Press 2 W...

Page 10: ...rely on accurate measurements www keysight com go quality Keysight Technologies Inc DEKRA Certified ISO 9001 2008 Quality Management System Keysight Channel Partners www keysight com find channelpartn...

Reviews: