1
Keysight 1.0 mm Precision Verification Kit
85059V
Operation and Service Guide
6
Replaceable Parts
Information in This Chapter
Chapter Six at-a-Glance
Replacing the Verification Data
The verification data contains unique performance data that applies to the
individual verification devices. No two devices have the same performance
data. It is not a trivial matter to replace lost or damaged data, so it is important
to make one or more backup copies.
If your verification data is lost or damaged, and you have no backup copies,
take one of the following actions:
— If recertification is not required in the near future.
Contact Keysight for replacement verification data - refer to
. Please specify the model and serial number of your
kit.
— If recertification will be required in the near future.
Keysight recommends that you have the verification kit recertified early.
New verification data will be generated during the recertification process.
Section Title
Summary of Content
“Replacing the Verification Data”
A discussion of what to do verification data
“Replaceable Parts” on page 6-2
Table of Replaceable Parts (See
IMPORTANT! Keysight’s data retention policy is to hold verification data for
18 months. If it has been >18 months since your last recertification, your
85059V kit must be recertified. If you lose this data, refer to
https://service.keysight.com/infoline/public/default.aspx
or contact
Summary of Contents for 85059V
Page 1: ...Keysight Technologies 85059V 1 0 mm Precision Verification Kit Operating and Service Manual ...
Page 44: ...5 6 Keysight 85059V Operation and Service Guide Troubleshooting Contacting Keysight 5 ...
Page 48: ...6 4 Keysight 85059V Operation and Service Guide Replaceable Parts Replaceable Parts 6 ...