2-2 Keysight 85059V Operation and Service Guide
Specifications
Environmental Requirements
2-
Environmental Requirements
Temperature—What To Watch Out For
Due to the thermal properties of the devices, electrical characteristics will
change with temperature. Therefore, the operating temperature is a critical
factor in their performance, and must be stable before use.
Performance verification and measurements of devices-under-test (DUTs) do
not need to be performed within the operating temperature range of the
calibration devices. However, the DUTs must be within the error-corrected
temperature range of the network analyzer (±1° C of the measurement
calibration temperature). For example, if the calibration is performed at +20° C,
the error-corrected temperature range is +19° to +21° C. It is then appropriate
to perform measurements and performance verifications even 19° is
outside the operating temperature range of the calibration devices.
Table 2-1
Environmental Requirements
Parameter
Required Values/Ranges
Temperature
Operating
a
a. The temperature range over which the calibration standards maintain conformance to their specifications.
+20
°
C to +26
°
C (+68
°
F to +79
°
F)
Storage
–40
°
C to +75
°
C
(–40
°
F to +167
°
F)
Error-corrected range
b
b. The allowable network analyzer ambient temperature drift during measurement calibration and during mea-
surements when the network analyzer error correction is turned on. Also, the range over which the network
analyzer maintains its specified performance while correction is turned on.
±
1
°
C (1.8
°
F) of measurement calibration temperature
Altitude
Operating
< 4,500 meters (ª15,000 feet)
Storage
< 15,000 meters (ª50,000 feet)
Relative humidity
Always non–condensing
Storage
0 to 90%
IMPORTANT!
Avoid unnecessary handling of the devices during use
because your fingers are a heat source.
Summary of Contents for 85059V
Page 1: ...Keysight Technologies 85059V 1 0 mm Precision Verification Kit Operating and Service Manual ...
Page 44: ...5 6 Keysight 85059V Operation and Service Guide Troubleshooting Contacting Keysight 5 ...
Page 48: ...6 4 Keysight 85059V Operation and Service Guide Replaceable Parts Replaceable Parts 6 ...
Page 52: ...85059V Index Index 4 ...