Use, Maintenance, and Care of the Devices
Cleaning Connectors
Cleaning Connectors
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial
connections.
1.
Use Compressed Air or Nitrogen
WARNING
Always use protective eyewear when using compressed air or nitrogen.
Use compressed air (or nitrogen) to loosen particles on the connector mating plane
surfaces.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an
effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less
than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of
compressed air can cause electrostatic effects when directed into a connector. These
electrostatic effects can damage the device. Refer to
chapter for additional information.
2.
Clean the Connector Threads
WARNING
Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly
closed container. It is extremely flammable. In case of fire, use alcohol
foam, dry chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with
eyes, skin, and clothing. It causes skin irritation, may cause eye damage,
and is harmful if swallowed or inhaled. It may be harmful if absorbed
through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal,
state, and local environmental regulations.
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or
stubborn contaminants on a connector that cannot be removed with compressed air or
nitrogen. Refer to
for a part numbers for cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you
reassemble or use it.
3.
Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to
Summary of Contents for 85058B
Page 1: ...Keysight Technologies 85058B E 1 85 mm Calibration Kits User s and Service Guide...
Page 2: ...85058B E i...
Page 7: ...1 General Information...
Page 28: ...1 22 85058B E General Information When to Calibrate...
Page 29: ...2 Specifications...
Page 38: ...2 10 85058B E Specifications Electrical Specifications...
Page 39: ...3 Use Maintenance and Care of the Devices...
Page 55: ...4 Performance Verification...
Page 58: ...4 4 85058B E Performance Verification Recertification...
Page 59: ...5 Troubleshooting...
Page 64: ...5 6 85058B E Troubleshooting Contacting Keysight...
Page 65: ...6 Replaceable Parts...
Page 73: ...A Standard Definitions...