1-
8
85058B/E
General Information
Achieving Specified Frequency Performance
Broadband SOLT Calibration Using the 85058E (Economy) Calibration Kit
The Broadband SOLT calibration is the only calibration technique available with the 85058E
(Economy) kit. You can calibrate from DC to 67 GHz using the standards in
Table
1-2
85058E (Economy) Broadband SOLT
a
a. SOLT = Short, Open, Load and Thru standards
Calibration Standards
Calibration Kit Label:
PNA: 85058E Data-Based
85058EP Polynomial
PNA File Names: 85058E Data-Based.ckt
85058EP Polynomial.ckt
Frequency
Range
Calibration
Technique
“Open”
Standard
“Short”
Standard
“Load”
Standard
“Thru”
Standard
(not used
for 1
port
cal)
“Isolation”
Standard
DC
67 GHz
SOLT
Open
(broadband)
Short 1
(broadband)
Load BB
(broadband)
Thru
Load BB