1
Introduction
24
Keysight 8480 Series Operating and Service Manual
Table 1-4
Typical measurement uncertainties of Calibration Factor (CF) data printed on the
power sensor
Frequency
25°C ± 3°C
8487A
8481B
8482B
8481H
8482H
8487D
8485D
8481D
100 kHz to 10 MHz
-
-
1.50
-
0.91
-
-
-
10 MHz to 30 MHz
-
1.48
1.44
0.86
0.81
-
-
0.77
30 MHz to 500 MHz
1.38
1.48
1.51
0.88
0.92
1.33
1.24
0.81
500 MHz to 1.2 GHz
1.34
1.48
1.48
0.87
0.88
1.35
1.26
0.81
1.2 GHz to 6 GHz
1.41
1.54
1.53
0.98
0.95
1.41
1.35
0.97
6 GHz to 14 GHz
1.59
1.71
-
1.36
-
1.62
1.63
1.20
14 GHz to 18 GHz
1.69
1.99
-
1.71
-
1.73
1.83
1.72
18 GHz to 26.5 GHz
2.23
-
-
-
-
2.25
2.45
-
26.5 GHz to 33 GHz
2.58
-
-
-
-
2.55
2.94
[a]
-
33 GHz to 34 GHz
2.73
-
-
-
-
3.08
-
-
34 GHz to 35 GHz
2.73
-
-
-
-
3.08
-
-
35 GHz to 40 GHz
2.73
-
-
-
-
3.08
-
-
40 GHz to 45 GHz
3.67
-
-
-
-
4.28
-
-
45 GHz to 50 GHz
4.33
-
-
-
-
4.72
-
-
[a] These uncertainties only apply to Option 033