General Information
2
Keysight 8480 Series Operating and Service Manual
49
12
The linearity system specification is calculated using the root sum of the
squares (RSS) method by considering the error caused from the reference
sensor used and the DUT error in the system at 50 MHz. The error specification
caused by the reference sensor and the DUT can be found in their respective
data sheet. The RSS error specification calculation is computed by using the
following equation:
The following example of the DUT is measured at 50Mhz, 20dBm:
– DUT used is 8481A
– Reference sensor used is N8481A
System linearity error specification
DUT error2 Ref sensor error2
+
±
=
System linearity error specification
(3%)2 (0.8%)2
+
±
3.10%
±
=
=