18 Keysight 16190B Performance Test Kit
Service
Calibration
3-
1600
0.90 %
+100 m
0.40 %
0.70 %
0.70 %
1800
0.90 %
+100 m
0.40 %
0.70 %
0.70 %
2000
1.0 %
+200 m
0.80 %
1.0 %
1.0 %
2200
1.0 %
+200 m
0.80%
1.0 %
No Calibration
2400
1.0 %
+200 m
0.80 %
1.0 %
1.0 %
2600
1.0 %
+200 m
0.80%
1.2 %
1.0 %
2800
1.0 %
+200 m
0.80 %
No Calibration
1.0 %
3000
1.0 %
+200 m
0.80 %
1.2 %
No Calibration
1. Calibration uncertainty when shipped from the factory
Table 3-2
Impedance Phase
(Admittance Phase
) Calibration Uncertainty
1
Frequency
[MHz]
50
Airline with
OPEN
Airline with
SHORT
1
1.0 mrad
1.6 mrad
1.7 mrad
10
1.0 mrad
1.6 mrad
1.7 mrad
100
1.0 mrad
1.6 mrad
1.8 mrad
200
2.0 mrad
2.0 mrad
2.0 mrad
300
2.0 mrad
2.0 mrad
2.0 mrad
500
3.0 mrad
3.0 mrad
3.0 mrad
600
4.0 mrad
5.0 mrad
4.0 mrad
800
4.0 mrad
5.0 mrad
No Calibration
1000
4.0 mrad
5.0 mrad
5.0 mrad
1300
4.0 mrad
No Calibration
6.0 mrad
1600
4.0 mrad
7.0 mrad
7.0 mrad
1800
4.0 mrad
7.0 mrad
7.0 mrad
2000
8.0 mrad
10 mrad
10 mrad
2200
8.0 mrad
10 mrad
No Calibration
Table 3-1
Impedance Magnitude |Z| (Admittance Magnitude |Y|) Calibration
Uncertainty
1
Frequency
[MHz]
OPEN
SHORT
50
Airline with
OPEN
Airline with
SHORT
Summary of Contents for 16190B