background image

 

 

Demonstrating multi-site configuration 

While the 4-port VNA is running, you can also create a different, fully-functioning VNA which would run at the same time 
as the 4-port VNA. The only practical limit to the number of VNA instances is the CPU speed of the host controller. The 
benefit of creating Multi-site VNAs is measurement throughput. You could test multiple devices at the same time. 

 

In this demo we use multiple M937XA to demonstrate multi-site configuration.  The key point of this demo is to show 
flexibility of the setting. 
 

 

The mainframe should have the latest version of the software for this demonstration. Refer to 

verify and install 

the software

 section for more detail. 

 

 

In the following instructions, two pairs of 4-port operation are configured. Use these instructions as a general 
guide if installing a different number of modules. 

 

Install Modules 

As seen in shown figure, install four modules. Install semi-rigid cables (M9370-20006 for LO In/Out, M9370-20018 for Trig 
In/Out, M9370-20019 for Ref In/Out) to connect each module.  Also, connect 50 ohm loads on LO In of the first module 
and LO Out of the second module for each pair of 4-port configuration M957XA. 

 

 

 
 

Summary of Contents for 1250-1666

Page 1: ...Demo Guide Keysight M937xA PXIe Vector Network Analyzers...

Page 2: ...iguration 13 Launch software 13 Setup measurement parameters 13 Perform calibration 15 Demonstrating multi site configuration 16 Install Modules 16 Using Soft Front Panel to demonstrate multi site con...

Page 3: ...literature no 5990 7632EN NET framework runtime components are installed by default with Windows Vista and Windows 7 Therefore you may not need this amount of available disk space o Hardware requirem...

Page 4: ...de Y1241A 1 Cable kit for multiport configuration Includes M9370 20006 Cable Assy RF LO Jumper M9370 20018 Cable Assy SMB Trigger Jumper M9370 20019 Cable Assy SMB Reference Jumper Y1281A 1 Accessory...

Page 5: ...ck the revision of the M937X It should be higher revision than the firmware on the web www keysight com find pxivna 3 Install software if the system does not have the latest software 1 Install the Key...

Page 6: ...le power is applied to the chassis capabilities Before installing or removing a module to from the chassis power off the chassis to prevent damage to the module 2 Insert M937X modules in a chassis wit...

Page 7: ...interface guides developers through module setup using a similar look and feel as the popular PNA Users can quickly configure the instrument parameters and perform calibrations Launch Soft Front Pane...

Page 8: ...Setup measurement parameters On the menu bar of M937X SPF Select Utility Quick start Select S Parameters Setup measurement parameters as shows in left for the demo DUT 8...

Page 9: ...Follow instruction to perform cal ECal modules are controlled and powered through a USB connection When you connect the module the type of module frequency range and connector type are automatically r...

Page 10: ...il In the following instructions eight modules are configured Use these instructions as a general guide if installing a different number of modules The selected modules in a Multiport VNA MUST be sequ...

Page 11: ...6 6 7 and 7 8 Press firmly to ensure a proper connection 3 Install a semi rigid cable M9370 20006 seen in below figure between the LO Out port and the LO In port on each of the module pairs 1 2 2 3 3...

Page 12: ...20019 seen in below figure between the Ref Out port and the Ref In port on each of the module pairs 1 2 2 3 3 4 4 5 5 6 6 7 and 7 8 Press firmly to ensure a proper connection 6 You have now completed...

Page 13: ...t M937XA modules to configure multiport VNA In this example 8 modules in the chassis in slots 2 3 4 5 6 7 8 9 are selected Click Run Setup measurement parameters In this step we setup the VNA to measu...

Page 14: ...dd other 15 reflection parameters From menu bar of the M937X SPF Trace Chan Trace New Trace Select S22 S33 S44 S55 S66 S77 S88 S99 S10_10 S11_11 S12_12 S13_S13 S14_S14 S15_S15 S16_S16 Now you have all...

Page 15: ...rder to enable option 551 capability in your multiport configuration you have to have at least 1 module of the M937X which has option 551 To check the option installed on the M937X go to the menu bar...

Page 16: ...this demo is to show flexibility of the setting The mainframe should have the latest version of the software for this demonstration Refer to verify and install the software section for more detail In...

Page 17: ...again click the Network Analyzer SFP shortcut which restarts the Launcher dialog The Launcher dialog shows the same selected slots as the previous time The selection is sticky to allow you to quickly...

Page 18: ...Now you have two M937X SPF in one system...

Page 19: ...eck the instrument error queue 5 Close the driver Examples are coded to initialize the driver in simulation mode and should compile and run as is without an instrument Example Visual Studio projects w...

Page 20: ...ut Select M937XA module to run Click Run Open example file From start menu Click Agilent IVI Drivers AgNA AgNA Examples Open Examples VS NET VB CreateMeasurement_VBExample Copy and rename CreateMeasur...

Page 21: ...mulate false In this example we use M937X installed in the system which IP address is 146 208 119 171 Use this example as the general guide for your program In this example we use HiSLIP protocol to a...

Page 22: ...On the Visual Studio run the program by pressing start button You will see the command prompt Different from previous run the simulate status should be False On the M937X SFP you will see S11 on Tr1...

Page 23: ...open VISA resource Most of the case the resourceDesc is not properly set on your program The resourceDesc should be the same as VISA address you can see on the Keysight Connection Expert Also do not f...

Page 24: ...ured with another analyzer the connector is terminated with a 50 ohm load If the module is configured with an analyzer in an adjacent slot the connector receives the LO signal from the configured anal...

Page 25: ...anager As a result of installing IO library and soft front panel required drivers are installed in the system and devices can be observed in the Windows Device Manager Below is example of two M937X mo...

Page 26: ...th the I O libraries NI VISA as well as individual PXI instrument drivers Details will vary depending on the PXI instrument provider but some of the installations also include tools similar to what th...

Page 27: ......

Page 28: ...X Configuration Guide http cp literature keysight com litweb pdf 5991 4885EN pdf Keysight IO Libraries Suite www keysight com find iosuite Application Note PXI Interoperability How to Achieve Multi Ve...

Page 29: ...This information is subject to change without notice Keysight Technologies 2014 Print Date Sept 2014 Supersedes None M9370 90006 www keysight com...