Specification
7
Function
Rated
Voltage (AC)
Trip Current
Trip Current
Duration
Accuracy
Function
Rated voltage
Measuring
Range
Accuracy
Voltage measurement-
100-250V
100-250V
3% rdg
RCD
x 1/2
230V+10%
-15% 50Hz
230V+10%
-15% 50Hz
Trip Current:
-10% +0% of
range at 230V
Trip Current:
+10% -0% of
range at 230V
Trip Current:
+10% -0% of
range at 230V
RCD
x 1
10/30/100/300/
500/1000 mA
10/30/100/300/
500/1000 mA
2000ms
2000ms
1000mA
@200ms
Trip Current:
±10% of
range at 230V
RCD
x 5
230V+10%
- 15% 50Hz
10 mA
30/100/300mA
(Note: on x5 range
max. current that can
be generated is 1A)
50ms
Trip
Time±
(1% rdg
+3dgt)
To prevent wrong connection of test leads and to maintain safety, the
dedicated terminals used for continuity and insulation tests are
automatically covered when using the terminals for loop impedance, PSC
and RCD tests.
Instrument dimensions
130 X 183 X 100mm
Instrument weight
1130g including batteries.
Reference conditions
Specifications are based on the following
conditions except where otherwise stated:-
1
Ambient temperature: 23±2˚C
2
Relative humidity 45% to 75%
3
Position: horizontal
4
AC power source 230V, 50Hz
5
DC power source: 12.0 V, ripple content
1% or less
6
Altitude up to 2000m