PLJ-N-BA-e-1333
62
15.4.4
Setting data output type: manual/automatic „P4.4 rEPL”
* Press the arrow keys (
), to select among the settings below:
rEPl no
Manual issue after pressing the
key.
rEPl yes
Automatic issue of first stable weighing value
Sequence of operations:
1. Taring
2. Place weight, issue of first stable weighing value
3. Renewed output only possible after weight was
removed. Condition: D/- 50 display steps from
zero
4. Place next weight.
Return to weighing mode:
Press the
TARE
-button repeatedly until the "
SAVE
"? inquiry appears.
Confirm inquiry by pressing the
-button to save changes and to
return to Weighing mode.
-or-
press the
TARE
-button to return to Weighing mode without saving.
Press repeatedly
button
Press the
button as often as
necessary until the required
setting appears:
confirm
Press repeatedly
button