PLJ-N-BA-e-1333
19
7.2.2 Adjustment test
– „tCAL“
Here, deviation from the last adjustment is determined. This is only a check, i.e. no
values are changed.
Operation:
Display:
Start balance by pressing
0.0000 g
Press
.
P1 CAL
Press
.
P1.1 iCAL
Press repeatedly
button. Ensure that there are no
objects on the weighing plate.
P1.3 tCAL
Press
, adjustment test is carried out automatically. The
result is displayed
CAL
diff
d 0,0042 g
Press
, balance returns to menu
P1.3 tCAL
Press
repeatedly until “
SAVE
“? appears
Press
to save the changes you made.
Press
, to cancel the changes you made. Afterwards the
balance automatically jumps back to weighing mode.
SAVE?
0.0000 g