PLJ-N-BA-e-1333
14
6.3 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be
the same as the local voltage.
Only use original KERN mains adapters. Using other makes requires consent by
KERN.
After connection to the power supply the balance will carry out a self-test. The
balance is ready for weighing when the weight display appears. If the display diverts
from zero, press the
TARE
-key.
6.4 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data
interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they
are ideally tuned to your balance.
6.5 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1).
During this warming up time the balance must be connected to the power supply.
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
6.5.1 Stability display
The appearance of the stability symbol
[
]
on the display indicates that the
balance is in a stable state. If the status is instable the [
] display disappears.
6.5.2 Balance zero display
If an exact zero reading is not displayed on the balance in spite of the weighing dish
being empty, press the
TARE
key and the balance will start resetting to zero
[
]
.
7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.