63
PBS/PBJ-BA-e-1314
16.5 Auto Print function
(models with non-verifiable setting only)
Using the Auto Print function ensures automatic data printing without necessity to
press the
button for each single measurement. If the function is activated, the
(Auto-Print) symbol is displayed.
It is possible to select from six types of automatic printing. Information concerning
setting the zero range can be found in chapter 11.7
Printing whilst loading
:
Select the menu element
42
in the menu overview.
Place or remove the sample when the displayed value lies within the zero range. The
data will be output automatically when the stabilisation symbol is illuminated and
the displayed positive value exceeds 5 times the zero range value. The next data
output will occur when the display is brought to the value within the zero range by
removing the sample or pressing the
TARE /
button.
Printing whilst loading and unloading
:
Select the menu element
43
in the menu overview.
Place or remove the sample when the displayed value lies within the zero range. The
data will be output automatically when the stabilisation symbol is illuminated and
the displayed positive or negative value exceeds 5 times the zero range value. The
next data output will occur when the display is brought to the value within the zero
range by removing the sample or pressing the
TARE /
button.
Printing whilst loading and at zero
:
Select the menu element
44
in the menu overview.
Place or remove the sample when the displayed value lies within the zero range. The
data will be output automatically when the stabilisation symbol is illuminated and
the displayed positive value exceeds 5 times the zero range value. Remove the
sample or press the
TARE /
button. The data will be output again when the displayed
value lies within the zero range and the stabilisation symbol is illuminated.
Printing whilst loading, unloading and at zero
:
Select the menu element
45
in the menu overview.
Place or remove the sample when the displayed value lies within the zero range. The
data will be output automatically when the stabilisation symbol is illuminated and
the displayed positive or negative value exceeds 5 times the zero range value.
Remove the sample or press the
TARE /
button. The data will be output again when
the displayed value lies within the zero range and the stabilisation symbol is
illuminated.