32
8.2.1 Adjustment test with external weight
•
Condition
: Menu setting „E-tESt“ / Menu element 4
•
At verified balances, the adjustment test is locked by a switch (except
accuracy class I). To disable the access lock, destroy the seal and actuate
the adjustment switch. Position of the adjustment switch see chap. 9
Attention
:
After destruction of the seal the balance must be re-verified by an authorised
agency and a new verification wire/seal mark fitted before it can be reused
for applications subject to verification.
In weighing mode press
. E-tESt is displayed.
If „E-tESt“ is not displayed, press
to return into weighing mode and
activate menu element 4, see menu overview.
Press
, checking is started. The displayed
weight value of the recommended adjustment weight
(see chapter 1) flashes.
If the weight value shall be changed, press
, the active digit flashes.
Carry out the
desired setting using navigation buttons (see chapter 3.1.1 „Numeric input“).
Carefully place adjusting weight in the centre of the
weighing plate ,press
.
Wait until the zero display flashes.