background image

8

8-2  The  details  of  transmitted  light  brightfield 

observation procedure
8-2-1    Set  the  main  switch  to  "I"(ON)  and  adjust  the 
intensity of illumination.
A.  Make  sure  that the light  intensity control  knob  is  in 
the  MIN(minimum  intensity)position  and  set  the  main 
switch to "I"(ON).
B.  Rotate  the  knob  toward  MAX(maximum  intensity)to 
increase the intensity and the illumination brightness.
C. When the microscope system is not be used for a long 
period,set the main switch to "O"(OFF).

8-2-2 Disengage the filter other than the frost filter

From the light path. (Please refer to 6-8.)

8-2-3    Select  the  light  path  with  the  light  path 

selector  lever. The  light  path  selector  lever  allows  for 
light  path  switching  between  the  observation  and  side 
port paths.

8-2-4    Place the specimen on the stage.

8-2-5    Engage a 10Xobjective in the light path.
8-2-6    Bring the specimen into focus.

(1) Rotation Direction of the Coarse/Fine Adjustment Knobs
            Rotating  the coarse  or  fine  adjustment  knob  toward 
the  front(in  the  direction  of  the  arrow)  raises  the  objective 
and toward the rear(opposite direction)lowers the objective.
(2) Adjusting the Coarse Adjustment Knob Tension
            Always  use  the  rotation  tension  adjustment  ring  to 
control the rotation tension of the coarse adjustment knob.
            The  tension  of  coarse  adjustment  knob  has  been 

pre-adjusted to optimum tension, but this can be changed as required. Turn the rotation tension 
adjustment ring in the direction of the arrow to increase the knob's tension and in the opposite 
direction to decrease it.
            If  the  objective  lowers  by  its  own  weight  or  the  focusing  obtained  with  the  fine 
adjustment knob is lost soon, the tension is set too low. In this case, turn the rotation tension 
adjustment ring in the opposite direction to the arrow to increase the tension.

8-2-1

8-2-3

Summary of Contents for OCO-2

Page 1: ...Operating Manual Inverted System Microscopes KERN OCO 2...

Page 2: ...f the text symbols 2 4 Precautions When Unpacking the Microscope 2 5 Module Nomenclature See diagram 1 3 6 Preparation for use 4 7 Preparations for the use of special devices 6 8 Operations 7 9 Troubl...

Page 3: ...ve precipitation and crystal structure analysis The microscopes are used in many practical areas as well such as bio medical science preventive medication environmental engineering food processing pha...

Page 4: ...the Revolving Nosepiece Never attempt to rotate the coarse or fine adjustment knob without removing the clamping rod Otherwise the focusing mechanism may be damaged 1 Loosen he screw of the clamping r...

Page 5: ...video port C mountdigitalport C mountvideo Camera three position slider reflector DSZ main body Analyzer UCIS objective PH UCIS objective specimen stage re ctangular stage With flexible right handle a...

Page 6: ...ed to contamination than the objectives of upright microscopes Therefore if there are empty positions 6 3 Attaching the fixator of the universal handwheel 6 3 1 This is an optional part which can be a...

Page 7: ...proper cord is provided If the power outlet is not grounded properly we can no longer warrant the electrical safety performance of the equipment 6 6 4 If the power cord or a connection cable comes in...

Page 8: ...ide port is not used attach the cap for protecting it from dust 6 10 Attaching the TV Observation System 6 10 1 This is an optional part for TV observation 6 10 2 Attaching the photographic eyepiece a...

Page 9: ...ght path Select the light path with the light path selector lever Place the specimen on the stage Engage a 10X objective in the light path Bring the specimen into focus Adjust the brightness Adjust th...

Page 10: ...10Xobjective in the light path 8 2 6 Bring the specimen into focus 1 Rotation Direction of the Coarse Fine Adjustment Knobs Rotating the coarse or fine adjustment knob toward the front in the directi...

Page 11: ...knob The stage up down movement using the fine adjustment knob is not limited When the pre focusing lever is locked the coarse adjustment stroke is limited by the mechanism and it cannot reach the pr...

Page 12: ...ver 8 3 4 Attach the phase contrast slider 8 3 5 Engage the objective matched with the phase contrast slider in the light path 8 3 6 Disengage the 10X eyepiece and insert the WY1 central telescope 8 3...

Page 13: ...ightness 8 3 13 Adjust the interpupillary distance and diopter 8 3 14 Engage the objective to be used in the light path and bring the specimen in focus 8 3 15 Adjust the aperture iris and field iris d...

Page 14: ...ed Reduce the filters to the minimum required b The bulb lights but the field of view is dark The stage central plate is engaged in the optical path Move the stage and place the specimen again Light p...

Page 15: ...asily or at all The pre focusing lever is locked Unlock it j The revolving nosepiece lowers by its own weight or defocusing occurs due to slipping of fine adjustment The tension adjustment ring is too...

Page 16: ...ure for cleaning for these will deform the electrically insulated section of the extremity Since solvents such as ether and alcohol are highly flammable they must be handled carefully Be sure to keep...

Reviews: