MLS-BA-defsi-0713
87
03 Calibration test
Here, deviation from the last adjustment is determined. This is only a check, i.e.
no values are changed.
Ö
Use the cursor (
►
) to select
“03 Calibr. test“
Setup
29.12.04
Setup
Ö
Push -button
Ensure that there are no objects on the weighing
plate.
Einstellungen
Calibration check
Cear the pan
[Enter]
Ö
Press -key; the display shows the measuring
value of the adjusting weight.
Einstellungen
Calibration check
Weight measuring
Einstellungen
Calibration check
Load weigh
[Enter]
t 50.0g
Ö
Now place the adjusting weight in the center of
the weighing plate, press the
key.
The result of the calibrating test is displayed.
Einstellungen
Calibration check
Weight measuring
Calibration
Check results
Cal.: 180.1354
Act.: 180.1438
Diff.: - 0.0084
Ö
Remove weight; actuate the
-key