KMF-TM-BA_IA-e-1310
29
Adjustment is required after entering configuration data!
Confirm with
,
[UnloAd]
appears
Ensure that there are no objects on the weighing pan.
Wait for stability display "STABLE", then confirm with
(example)
The size of the currently set adjustment weight will be
displayed.
To change, select the digit to be altered by
, and the
numerical value by
.
Confirm with
,
[LoAd]
appears.
Place the adjustment weight carefully
Wait until stability display „STABLE“ appears
Confirm with
[PASS]
is displayed.
After the adjustment the balance will carry out a self-test. Remove
adjusting weight
during
selftest, balance will return into weighing
mode automatically.
An adjusting error or incorrect adjusting weight will be indicated
by the error message; repeat adjustment procedure.
An adjusting error or incorrect adjustment weight will generate an
error message („Err 4“), repeat the adjustment process.