20
KMF-TM-BA_IA-e-1310
11 Instant help
In case of a fault in the program sequence, the balance should be shortly switched
off. The weighing process must then be restarted from the beginning.
Failure:
Possible causes:
The displayed weight
does not glow.
The display unit is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
Rechargeable battery inserted incorrectly or empty.
No rechargeable battery inserted.
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
The load cell has contact with foreign objects
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is
obviously incorrect
The display of the balance is not at zero.
Adjustment is no longer correct.
Great fluctuations in temperature.
The weighing system is not on even surface.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
Summary of Contents for KMF-TM
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......