SYMPTOMS
CAUSES & REMEDY
Corner Error is too much.
1. Mechanical unit is defective. Leaf spring is bent
or twisted.
2. Pan base touch other parts.
Display suddenly disappeared.
1. HT-DP board is defective.
2. AC adaptor is defective.
or
appears.
1. HT-DP board is defective.
2. Coefficient memories (address data) have
changed by noises or static electricity. Re-set the
address data and/or adjust the linearity.
appears.
1. The weight of reference sample is too light (%
mode, counting mode).
appears.
1. Wrong external weight is used in span calibration
(external weight must be more than 50% of
scale capacity).
appears.
1. Span error exceeds 1% of scale capacity in
span calibration. Check the external weight.
appears.
1. During calibration there was a weight on the
weighing plate. Remove it from the pan.
appears.
1. Span error exceeds 1% of capacity in automatic
span calibration.
appears.
1. Interval time is not properly set.
appears.
1. Internal clock is defective. Check internal clock
and replace HT-DP board, if necessary.
[
r - Err
] appears.
1. The weight error exceeds ±100.00mg. Re-set the
weight error within ±100.00mg
[
A - Err
] appears.
1. Internal calibration automatics defective
appears.
1. Wrong operation in Cumulate Function or Net
Addition Function.
9
Summary of Contents for KERN AEJ
Page 5: ...3 Electric Electronic Construction 3 1 Block Diagram Calibration Weight Unit AEJ only PRINT 5 ...
Page 6: ...3 2 Whole Wiring Calibration Weight Unit AEJ only PRINT 6 ...
Page 7: ...4 Troubleshooting 4 1 Troubleshooting Procedure 7 ...
Page 11: ...4 4 Check of Electric Electronic Parts 1 Check of HT DP board 11 ...