5 Adjustment and Setting
5.1 Span Calibration (Adjustment with internal weight – AEJ)
Procedure for adjustment (according to the operating instructions):
Observe stable environment conditions. A warming-up time of approx. 8 hours for
stabilisation is necessary.
Function menu 1 setting:
[
]
x CA. 1
Ö
Press the
CAL
key
AWT. CAL
CAL
PWSH C
CAL
Ö
Press the
CAL
-key again and adjustment will take
place automatically.
AWT. CAL
CAL
CH. O
1
CAL
CH. F.S.
1
CAL
Ö
The process of adjustment is completed.
Adjustment protocols will be issued automatically
if you connect an optional printer, see chpt. 15
“printer mode”. Whilst printing is in progress [busy
] will be displayed.
EnD
CAL
Ö
The balance returns automatically into weighing
mode.
13