![KERN ILJ 25K-4CM Operating Manual Download Page 63](http://html1.mh-extra.com/html/kern/ilj-25k-4cm/ilj-25k-4cm_operating-manual_1957285063.webp)
ILJ-C/ILS-C-BA-e-1320
63
Return to weighing mode
Press
MODE
and F1 WEIGHING will be displayed. Confirm by pressing
and
the weighing scale will change to weighing mode.
Determining reference weight by weighing
The ENTER REFERENCE WEIGHT function is allocated to the
F2
-key.
Go to menu and select P2.6 DEVIATION
Use the arrow keys to select P2.6.5 SHORTCUTS
Press
F2
, P2.6.5.1 will be displayed along with the previous function for the
F1
-key.
Press
F1
and entry for the
F2
-key will take place
Press
F2.
Use the arrow keys to select ENTER REFERENCE WEIGHT and
Confirm by
; this function has now been allocated to
F2
.
In weighing mode:
This completes setting the operating
mode weighing by percentage.
The previously entered reference
weight will be displayed on the info
line.
Use the arrow keys to enter the
reference weight equalling 100%
and confirm by pressing
PRINT.
(example)
Example: 2 kg
The reference weight of 2 kg has now been saved
as 100% value.
Remove reference weight and place load. The weight will be displayed in
percent according to reference weight.