![KERN ILJ 25K-4CM Operating Manual Download Page 19](http://html1.mh-extra.com/html/kern/ilj-25k-4cm/ilj-25k-4cm_operating-manual_1957285019.webp)
ILJ-C/ILS-C-BA-e-1320
19
7.2 Menu settings
Menu item „P1 ADJUSTMENT“
[Adjustment]
ILJ-C
ILS-C
P1.1 INT.JUST.
-
Internal adjustment, ILJ-C only
-
P1.2
EXT.JUST.
External adjustment *
-
P1.3
BE.JUST.
User adjustment
P1.4 TEST JUST.
-
Adjustment test
* = Locked in weighing scales with type approval certificate [Er 9 lock].
7.2.1 P1.1 INT.JUST.
– Internal adjustment (models ILJ-C only)
For automatic internal adjustment see chap. 7.1
Starting internal adjustment manually
Press the
CAL
-
key or invoke menu “P1.1 INT.JUST.“
Press
or
Internal adjustment will be taking place.
Ensure that there are no objects on the weighing plate.
You can cancel the adjustment by pressing the
UNIT
-key.
After successful adjustment the weighing scale will return automatically to
the menu item. To return to weighing mode, press
UNIT
.