12.3 Auto Zero
This function is used to tare small variations in weight automatically.
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the “stability
comp
ensation“. (e.g. slow flow of liquids from a container placed on the balance,
evaporating processes).
When apportioning involves small variations of weight, it is advisable to switch off this
function.
Example: Menu item P2.2.1 invoke READING, press
2x
AUTOZERO YES will be displayed
Yes will be flashing
Use the arrow keys to change between
YES and NO
YES -
– Auto zero enabled
NO
– Auto zero disabled
(example)
To save desired setting, press
The setting is saved in such a way that
the weighing scale returns to the menu.
To return to weighing mode, press
UNIT