4 Error messages
Non-approval models
:
E1
⇒
ZERO over 300,000 internal value.
E2
⇒
ZERO below 10,000 internal value.
Approval models
:
E1
⇒
ZERO over the zero position in weight calib 10% of full scale.
E2
⇒
ZERO over the zero position in weight calibration - 10% of full scale.
All models
:
E4
⇒
ZERO is unstable.
E5
⇒
Turning on machine core value is under 0.
E6
⇒
Core value is over 250,000.
E7
⇒
Core value is below 120,000.
- - OL -
⇒
The weighing range is exceeded.
NOTES:
•
Non-approval models (CXB)
: After E1, E2 or E4 is displayed press
to continue.
•
Approval models (CXB-M)
: After E1, E2 or E4 appear, set the jumper SWA1
on main board to “ADJ” position and press
to return the counting
mode.
Then press
+
+
at the same time followed by pressing
to access calibration (see chap. 7.3).