2 Introduction
1.
There are 3 ways to rectify the linearity for setting the temperature calibration,
which can make the scale more accurate.
n
Inside air temperature (normal temperature)
o
High temperature
p
Low temperature
♦
It is adequate to only do the inside air temperature calibration. Upon the
customer’s request, we also offer the high temperature and low temperature
calibration as well.
2.
Sampling speed is also depended on the user’s need (01~05).
3.
Linearity calibration, capacity setting and temperature calibration can be set
separately.
4.
Weighing unit:
g
or
kg
(depending on model)
5.
Tare range is full range tare.
3 General Information
1. The newly manufactured scale without passing the calibration and setting process
will automatically enter the linearity calibration mode. After the linearity calibration
procedure is done, the program will automatically enter the Specification Setting
Mode.
2.
Zero Range
The internal value should be in between 120,000~250,000 with the platter on
and without any load
(check
Æ
see chapter 10 “Spec. calibration”)
3. Maximum Capacity
The internal value should be over 300,000 when the scale is fully loaded.
4. When rubbing the tetragon, the inaccuracy value should be smaller than 10
divisions.
5. Tare range is full range tare.
4