KERN CXB-SH-e-0913 Service Manual Download Page 15

10 Spec. calibration 

 

 

Before calibrating, please adjust the mini jumper SWA1 to “ADJ” position.  

 After finish the calibration, please re-adjust the mini jumper SWA1 to “LOCK” 

position.

 

 

   

n

  Turn on the machine, after the scale resets back to zero,  

then press 

 

key +  

1

  key + 

 7

  key (at the same time) to enter the spec. 

calibration mode. 

         

Weight

 

    

Piece Weight

 

Pcs

 

0 3 0 0 1 

    0 2 1 X X 

 

2 0 2 0 5 3 

 

                      

program version               keyboard default value               core value 

                                                  

  

       (to check keyboard function)

    

(internal value of zero)

 

see chapter 3 

“General Information“

 

o

 Press 

 key 

     

     Weight

 

     

Piece Weight

 

Pcs

 

Ñ

0

Ò

0 0 0 0 

     0 0 0 0 0  

 

0 0 0 0 0 1 

 

       

full capacity setting

            

preset calibrated value  

           

refer to “Approx. 1” 

      key in “ full scale

9d ”                suggest key in “full scale” 

  (overload function max.+ 9d) 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Approx. 1

:  

0 0 0 0 0 1

    

 

                              a    b   c   d   e

           

          

 

a

 

 input the division 

 1  = e.g. 1g / 10g     2  = e.g. 0,2g / 2g     5  = e.g. 0,5g / 5g

 

b

 

 input the decimal point  

 0  = e.g. 1g  

  1  = e.g. 0,2g  

  

    2  = e.g. 0,01kg    3  = e.g. 0,001kg 

c

 

 input the unit of the unit weight 0, 1, 2, 8, 9

 

(full capacity are 10

n

 of division) 

 

 

 

d

 

 define the keyboard function of 

 

;

  kg/lb

  ;  

PST CE

  

keys 

(use always  1 )

 

       

 

 

0 = PST CE, unit weight in kg 

1 = 10 preset, unit weight in kg 

(OIML approval model) 

2 = kg/lb (1

st

 unit = kg, 2

nd

 unit = lb)   

 3 = 10 preset, kg/lb (1

st

 unit kg, 2

nd

 unit lb) 

4 = PST CE, lb unit only                                5 = 10 preset, lb unit only 
6 = lb/kg (1

st

 unit lb, 2

nd

 unit kg)                     7 = 10 preset, lb/kg (1

st

 unit lb, 2

nd

 unit kg) 

0 = 10

0    

(x 1)       » CXB 15K1 / CXB 30K2

 

1 = 10

1    

(x 10) 

2 = 10

2    

(x 100) 

8 = 10

-2

   

(x 0.01) 

9 = 10

-1   

(x 0.1)    » CXB 3K0.2 / CXB 6K0.5 

 

Example:  CXB 3K0.2   

a

=  2. 

value of 

a

 x 10

= Readout          2 x 10

-1 

= 0,2g         

c

= 9  

CXB-SH-e-0913  

15 

Summary of Contents for CXB-SH-e-0913

Page 1: ...n GmbH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Service manual Counting scales KERN CXB Version 1 3 4 2009 GB CXB SH...

Page 2: ...__ 9 7 2 Back to the original value of calibration_______________________________ 9 7 3 Calibration of approval models CXB M _____________________________ 10 8 Service mode access ____________________...

Page 3: ...ul repair the balance will have to be reverified before it can be used again in a statutorily regulated field Das Ger t darf nur von geschultem oder beruflich ausgebildetem Fachpersonal z B eichrechtl...

Page 4: ...tely 4 Weighing unit g or kg depending on model 5 Tare range is full range tare 3 General Information 1 The newly manufactured scale without passing the calibration and setting process will automatica...

Page 5: ...s E4 ZERO is unstable E5 Turning on machine core value is under 0 E6 Core value is over 250 000 E7 Core value is below 120 000 OL The weighing range is exceeded NOTES Non approval models CXB After E1...

Page 6: ...5 Explanation of display Weight Reference weight Loading status display Pieces 6 CXB SH e 0913...

Page 7: ...ered by user of calculated by balance Overlay W indicates Placed number of pieces insufficient for reference calculation Placed reference weight insufficient for reference calculation Display quantity...

Page 8: ...e weights Addition in total memory Call up total memory Delete summation memory Enter reference weight through weighing Display reference weight stored last Enter target number of pieces Numeric entry...

Page 9: ...ill be carried out automatically after dead stop control Remove calibration weight during count down to zero The balance returns automatically into weighing mode An error message will appear on the di...

Page 10: ...t is flashing on display Carefully place calibration weight in the centre of the weighing plate Press key Calibration will be carried out automatically after dead stop control Remove calibration weigh...

Page 11: ...ation routines SWA 1 jumper is located adjacent to the RF screening cover of the main PCB Place your balance upside down and remove seal cap For calibration and configuration routines the jumper SWA1...

Page 12: ...12kg 12kg 3kg 36kg are 1 2 times of 30kg n Turn on the machine after the scale resets back to zero then press key 1 key 6 key at the same time to enter the linearity calibration mode Please press 9 ke...

Page 13: ...Weight Pcs 0 3 1 XXXXXX input n value e g n 2 internal value of the phase r Add the 3rd weight mass e g 12kg in sum 21kg and input n value e g n 4 then press key Weight Piece Weight Pcs 0 4 1 XXXXXX...

Page 14: ...eight Piece Weight Pcs 0 6 1 XXXXXX input n value e g n 1 internal value of the phase u Press key CONFIRM key the buzzer beeps once Weight Piece Weight Pcs S Y S U P v Remove all the weight mass the s...

Page 15: ...ey in full scale 9d suggest key in full scale overload function max 9d Approx 1 0 0 0 0 0 1 a b c d e a input the division 1 e g 1g 10g 2 e g 0 2g 2g 5 e g 0 5g 5g b input the decimal point 0 e g 1g 1...

Page 16: ...lue is the reference for the temperature calibration procedure The scale will be self tested and reset back to zero Press key 1 key 5 key then press key to enter the temperature calibration mode see c...

Page 17: ...t Pcs 3 0 0 1 8 3 0 0 0 0 1 0 0 1 1 1 Weight Piece Weight Pcs 6 0 0 4 5 6 0 0 0 0 2 0 0 1 1 1 CXB 3K1M CXB 6K2M Weighing unit kg Weight Piece Weight Pcs 1 5 0 0 9 1 5 0 0 0 1 3 0 1 0 1 Weight Piece We...

Page 18: ...al value of full capacity p Put the full capacity weight on the platter press key to calibrate full capacity value q After scale is stable it will reset back to zero automatically remove the weight ma...

Page 19: ...X XXXXXX full capacity value high temperature C internal value of full capacity p Put the full capacity weight on the platter press key to calibrate full capacity value q After scale is stable it will...

Page 20: ...e full capacity value q After scale is stable it will reset back to zero automatically remove the weight mass from the platter before it resets back to zero then the procedures finished Put the weight...

Page 21: ...12 Spare parts CXB SH e 0913 21...

Page 22: ...8 Power nameplate 230V AC 1 19 Power sticker 1 20 Dust cover 1 21 FD aluminum support 1 22 FD aluminum bracket 1 23 Battery 6V 4AH 1 24 X2 capacitor 0 1 F 275V P 15 1 25 Wire Fixer 1 26 Front panel 1...

Page 23: ...43 Upper housing 1 44 Lower housing 1 45 AE Sealing 1 46 Round head screw M4 8 6 47 Multi wire 150mm 1 48 Alloy hexagon screw M6 16 4 49 Power cable 1 50 Capacity sticker 1 51 Spring 1 52 Transformer...

Reviews: