The display will show “
Pin 2
”
Enter the new Pin number a second time. Press the
[Tare]
key to accept.
The display will show “
dOnE
” if the number was accepted or “
FAiL
” if the numbers were
not the same.
The scale returns to the parameter menu.
4.2.10
F10
- LVD MODE
It stands for Low Voltage Detect mode. The Low voltage detection can be enabled or
disabled.
To enter this parameter, press the
[Tare]
key when “
F10
” is shown.
The display will show if the LVD is “
on
” or “
oFF
”.
Press the
key to change the value.
Press
[Tare]
to accept “
on
”.
NOTE: Only “
LVD
” “
on
” is allowed for the CPB-M scales.
Turn the power off, remove the jumper and restart the scale. The new settings are now
in operation.
4.3 WEIGHING UNITS
[
kg / g
]
The current software of CPB was standardised to weigh in [kg].
There is a function to enable the scale to weigh in [g].
The scale has to be opened and a connecting jumper has to be fitted to “JTAG” pins 1
and 2 on the main board to enable the [g] display.
NOTE: The selection between both units is not possible for the CPB-M scales.
CPB-SH-e-0721