Error Codes
Error codes are shown on the main display for approx. 2 seconds, after which the program
automatically returns to the weighing mode.
Display
Cause
Solution
No segments appear
No AC power is available
Check the AC power supply
on the display
The AC adapter is not
Plug the AC adapter
plugged in
into an electrical outlet
Automatic shutoff is configured
Turn on the balance
Non-rechargeable/
Replace or recharge batteries
rechargeable
using external charger
batteries are drained
h
The load exceeds
Unload the weighing pan
the balance capacity
l
Something is touching
Move the object that is touching
the weighing pan
the weighing pan away
e 01
Data output is not compatible
Change the setting to the compatible
with the input format
one in the menu
e 02
Calibration/adjustment
Calibrate/adjust only when zero
condition not met, e.g.:
is displayed
– balance not zeroed
Press
(
to zero balance display
– balance is loaded
Unload balance
e 08
The balance was zeroed
Only zero the balance when in the
outside the zero range
allowable zero range of ± 2% of
the maximum capacity
e 09
Taring not possible
Press
(
when the gross value
≤
zero
to zero the balance
e 22
The weight is too light
Increase the reference
or there is no sample
sample quantity
on the weighing pan
e 30
The interface port
Set code to reset menu or contact
for printer output is blocked
your local KERN Service Center
The weight readout
Unstable ambient
Set up the balance in
changes constantly
conditions (too much
a different area or
vibration or the balance)
change the menu setting
is exposed to a draft
to adapt the balance
A foreign object is caught
Remove the foreign
between the weighing pan
object
and the balance frame
The weight readout
The balance was not
Calibrate/adjust the balance;
is obviously wrong
calibrated/adjusted; the
zero before weighing
balance was not
zeroed before weighing
If any other errors occur, please contact your local KERN service center.